User Manual for PC-80D Easy ECG Monitor
23
10.3 Classification
1. The type of protection against electric shock: Internally powered equipment
2. The degree of protection against electric shock: Type BF applied part
3. The degree of protection against harmful ingress of liquid: Ordinary equipment without protection against ingress of liquid.
4. Electro-magnetic Compatibility: Group I, Class B
10.4 Display
1. LCD type: Color LCD
2. Display area: 70.08mm x 52.56mm.
10.5 Data Memory
Stored data records can be uploaded to the computer by data cable provided by the manufacturer.
10.6 Overall Dimension and Weight
Dimension: (L)130mm×(W)98mm×(H)36 mm
Weight400gnot including batteries
10.7 Additional declaration
1) Electro-surgical interference suppression: Without the function of Electro-surgical interference suppression.
2) Defibrillator discharge protection: Without protection against defibrillator discharge.
3) Respiration or lead off detector and active noise suppressor: The auxiliary leakage current is less than 0.1uA.
4) Per minutes HR average calculation: the latest 8 times R-R intervals falling on the range of heartbeat average.
5) Display updates speed: 1 time/ second.
6) Battery life: Four AA alkaline batteries can be used for more than 2 hours measurement, or Four AA NiMH rechargeable battery (Capability at least
2200mAH) can be used for more than 8 hours.
Maximal input DC offset voltage: ±300mV.
Vue de la page 27